区域折射型多焦点人工晶状体与非球面单焦点人工晶状体植入术后患者客观视觉质量的短期对比研究
崔嵬1, 杨松霖1,(), 晏晓明1   
  1. 1. 100000 北京大学第一医院眼科
  • 收稿日期:2018-06-05 出版日期:2018-06-28
  • 通信作者: 杨松霖
  • 基金资助:
    国家自然科学基金(11372011)

Short-term objective visual quality comparison between after implantation of segmental refractive multifocal intraocular lens and aspheric-surface monofocal intraocular lens

Wei Cui1, Songlin Yang1,(), Xiaoming Yan1   

  1. 1. Department of Ophthalmology, Peking University First Hospital, Beijing 100000, China
  • Received:2018-06-05 Published:2018-06-28
  • Corresponding author: Songlin Yang
  • About author:
    Corresponding author: Yang Songlin, Email:
引用本文:

崔嵬, 杨松霖, 晏晓明. 区域折射型多焦点人工晶状体与非球面单焦点人工晶状体植入术后患者客观视觉质量的短期对比研究[J]. 中华眼科医学杂志(电子版), 2018, 08(03): 103-109.

Wei Cui, Songlin Yang, Xiaoming Yan. Short-term objective visual quality comparison between after implantation of segmental refractive multifocal intraocular lens and aspheric-surface monofocal intraocular lens[J]. Chinese Journal of Ophthalmologic Medicine(Electronic Edition), 2018, 08(03): 103-109.

目的

对比区域折射型多焦点人工晶状体(MIOL)SBL-3与非球面单焦点人工晶状体Akreos Adapt AO植入术后短期内患者的客观视觉质量。

方法

选取2017年6月至2018年1月在北京大学第一医院眼科诊断为单纯老年性白内障,并行白内障超声乳化吸除联合人工晶状体植入术的患者26例(35只眼)为研究对象。其中,植入SBL-3晶状体有15只眼,植入AO晶状体有20只眼。术后1个月检查SBL-3组及AO组患者术眼的裸眼远视力(UDVA)、裸眼中视力(UIVA)、裸眼近视力(UNVA)、矫正远视力(CDVA)、矫正近视力(CNVA)、最佳矫正远视力下的中视力(DCIVA)及最佳矫正远视力下的近视力(DCNVA),并使用视觉质量分析仪评估两组患者瞳孔直径4 mm与6 mm的全眼、角膜及眼内的总高阶像差、球差及点扩散函数。术后1个月,绘制SBL-3组患者术眼的离焦曲线。两组患者术眼的UDVA、CDVA、CNVA、UIVA、UNVA、DCIVA、DCNVA及视觉质量以均数±标准差(±s)表示,行非参数Mann-Whimey U检验比较。

结果

术后1个月时,SBL-3组及AO组患者术眼的UDVA、CDVA及CNVA差异均无统计学意义(Z=-0.99,0,-1.84;P>0.05);SBL-3组患者术眼的UIVA、UNVA、DCIVA及DCNVA分别为0.32±0.08、0.44±0.18、0.30±0.10及0.33±0.11,均优于AO组的0.25±0.10、0.27±0.09、0.20±0.07及0.21±0.10,差异均有统计学意义(Z=-2.26,-3.13,-1.84,-2.17;P<0.05)。视觉质量评估显示,瞳孔直径为4 mm的两组患者术眼球差的差异无统计学意义(Z=-1.50,-1.04,-0.82;P>0.05);瞳孔直径为6 mm的两组患者术眼球差的差异无统计学意义(Z=-0.12,-1.93,-1.16;P>0.05)。SBL-3组瞳孔直径4 mm与6 mm的全眼总高阶像差分别为0.45±0.12与1.41±0.21,高于AO组的0.26±0.12与0.82±0.13,差异均有统计学意义(Z=-3.45,-3.24;P<0.05)。SBL-3组患者术眼的点扩散函数为0.05±0.02,低于AO组的0.11±0.07,差异有统计学意义(Z=-2.60,P<0.05)。SBL-3组患者术眼的离焦曲线显示波峰出现在0 D处,在-1.0 D至-3.0 D范围内的曲线较平稳,下降不明显;在-3.0 D至-5.0 D范围内的曲线快速下降。

结论

Adapt AO晶状体术后客观视觉质量优于SBL-3晶状体,但SBL-3晶状体在中、近视力的应用效果优于Adapt AO晶状体。SBL-3晶状体的离焦曲线波峰出现在0 D处,在-1.0 D至-3.0 D范围内的曲线较平稳,下降不明显。本研究还需扩大样本量并长期观察其效果及安全性。

Objective

To compare the short-term postoperative quality of vision of segmental refractive multifocal intraocular lens (MIOL) SBL-3 and aspheric-surface monofocal intraocular lens Akreos Adapt AO.

Methods

26 cases of age-related cataract (35 eyes) after phacoemulsification cataract extraction combined with IOL implantation were collected from June of 2017 to January of 2018 in the Peking University First Hospital. There were 15 eyes in SBL-3 group, and 20 eyes in AO group according to the different types of IOL implanted. At 1 month after operation the following parameters of the two groups were assessed: uncorrected distant visual acuity(UDVA), uncorrected intermediate visual acuity(UIVA), uncorrected near visual acuity(UNVA), corrected distant visual acuity(CDVA), corrected near visual acuity (CNVA), distance-corrected intermediate visual acuity(DCIVA), distance-corrected near visual acuity(DCNVA). And analyze total higher order aberrations (ocular, corneal, internal), spherical aberration, point spread function for 4mm and 6mm optical zone. Defocusing curve of SBL-3 group was also drawn at 1 month postoperatively. The two groups of patients, UDVA, CDVA, CNVA, UIVA, UNVA, DCIVA, DCNVA and visual quality were expressed as mean standard deviation (±s), and were compared with non parametric Mann-Whimey U test.

Results

At 1 month after operation, the UIVA, UNVA, DCIVA and DCNVA in SBL-3 group [(0.32±0.08), (0.44±0.18), (0.30±0.10), (0.33±0.11)] were all better than that of the eyes in AO group [(0.25±0.10), (0.27±0.09), (0.20±0.07), (0.21±0.10)] (Z=-2.26, -3.13, -1.84, -2.17; P>0.05). But there were no statistical differences between two groups in UDVA, CDVA and CNVA(Z=-0.99, 0, -1.84; P>0.05). The visual quality assessment showed that the difference of eyeball difference between the two groups with the pupil diameter of 4 mm was not statistically significant (Z=-1.50, -1.04, -0.82; P>0.05), and the difference of eyeball difference between the two groups with 6 mm pupil diameter was not statistically significant (Z=-0.12, -1.93, -1.16; P>0.05). The total higher order aberrations of ocular for 4mm and 6mm optical zone in SBL-3 group [(0.45±0.12), (1.41±0.21)] were better than that in AO group [(0.26±0.12), (0.82±0.13)] (Z=-3.45, -3.24; P<0.05). But there were no statistical differences between two groups in spherical aberration. The point spread function in SBL-3 group (0.05±0.02) was lower than that in AO group (0.11±0.07) (Z=-2.60, P<0.05). There is a peak appeared at 0 D and the defocusing curve was smooth within -1.0 D and -3.0 D but was steep within -3.0 D and -5.0 D.

Conclusions

The objective visual quality of Adapt AO is better than that of SBL-3 lens, but the application of SBL-3 lens in medium and near vision is better than that of Adapt AO lens. The peak of the defocus curve of SBL-3 lens appeared at 0 D, and the curve in -1.0 D to -3.0 D was stable, and the decrease was not obvious. This study also needs to expand the sample size and observe its efficacy and safety in the long term.

表1 SBL-3组与AO组患者术眼术后视力情况的比较(±s)
表2 瞳孔直径为4 mm的SBL-3组与AO组患者术眼术后视觉质量的比较(±s)
表3 瞳孔直径为6 mm的SBL-3组与AO组患者术眼术后视觉质量的比较(±s)
图1 SBL-3组患者术眼术后1个月的离焦曲线图
[1]
Maxwell A, Holland E, Cibik L, et al. Clinical and patient-reported outcomes of bilateral implantation of a +2.5 diopter multifocal intraocular lens[J]. J Cataract Refract Surg, 2017, 43(1): 29-41.
[2]
王蒙蒙,张凤妍,崔琨明,等. 衍射型非球面多焦点人工晶状体植入术后视觉质量评价[J]. 中国实用眼科杂志,2013,31(5):561-564.
[3]
Mojzis P, Piñero DP, Ctvrteckova V, et al. Analysis of internal astigmatism and higher order aberrations in eyes implanted with a new diffractive multifocal toric intraocular lens[J]. Graefes Arch Clin Exp Ophthalmol, 2013, 251(1): 341-348.
[4]
Yoshino M, Minami K, Hirasawa M, et al. Clinical Results of Diffractive Multifocal Intraocular Lens Implantation after Laser In Situ Keratomileusis[J]. Nippon Ganka Gakkai Zasshi, 2015, 119(9): 613-618.
[5]
De Silva SR, Evans JR, Kirthi V, et al. Multifocal versus monofocal intraocular lenses after cataract extraction[J]. Cochrane Database of Systematic Reviews, 2016, 12: CD003169.
[6]
Chen T, Fang Y, Lin H, et al. Objective and subjective visual quality after implantation of all optic zone diffractive multifocal intraocular lenses: a prospective, case-control observational study[J]. Br J Ophthalmol, 2016, 100(11): 1530-1535.
[7]
Ramón ML, Piñero DP, Pérez-Cambrodí RJ. Correlation of visual performance with quality of life and intraocular aberrometric profile in patients implanted with rotationally asymmetric multifocal IOLs[J]. J Refract Surg, 2012, 28(2): 93-99.
[8]
Cillino S, Casuccio A, Di PF, et al. One-year outcomes with new-generation multifocal intraocular lenses[J]. Ophthalmology, 2008, 115(9): 1508-1516.
[9]
Javitt JC, Steinert RF. Cataract extraction with multifocal intraocular lens implantation: A multinational clinical trial evaluating clinical, functional, and quality-of-life outcomes[J]. Ophthalmology, 2000, 107(11): 2040-2048.
[10]
刘芳,王方,苏彦,等. AcrySof Restor多焦点人工晶状体SN6AD1和SN6AD3型植入术后患者视功能及生存质量研究[J]. 中国实用眼科杂志,2013,31(7):837-840.
[11]
Vries NED, Webers CAB, Touwslager WRH, et al. Dissatisfaction after implantation of multifocal intraocular lenses[J]. J Cataract Refract Surg, 2011, 37(5): 859-865.
[12]
Woodward MA, Randleman JB, Stulting RD. Dissatisfaction after multifocal intraocular lens implantation[J]. J Cataract Refract Surg, 2009, 35(6): 992-997.
[13]
Lafuma A, Khoshnood B, Berdeaux G. Comparison of outcomes with multifocal intraocular lenses: a meta-analysis[J]. Clinical Ophthalmology, 2011, 5(default): 45-56.
[14]
De Vries NE, Nuijts RM. Multifocal intraocular lenses in cataract surgery: literature review of benefits and side effects[J]. J Cataract Refract Surg, 2013, 39(2): 268-278.
[15]
Velika V, Hejsek L, Raiskup F. Clinical Results of the Implantation of Two Types of Multifocal Rotational Asymmetric Intraocular Lenses[J]. Cesk Slov Oftalmol, 2017, 73(1): 3-12.
[16]
Shah S, Perismartinez C, Reinhard T, et al. Visual Outcomes After Cataract Surgery: Multifocal Versus Monofocal Intraocular Lenses[J]. J Refract Surg, 2015, 31(10): 658-666.
[17]
Zelichowska B, Rekas M, Stankiewicz A, et al. Apodized diffractive versus refractive multifocal intraocular lenses: optical and visual evaluation[J]. J Cataract Refract Surg, 2008, 34(12): 2036-2042.
[18]
Venter JA, Barclay D, Pelouskova M, et al. Initial experience with a new refractive rotationally asymmetric multifocal intraocular lens[J]. J Refract Surg, 2014, 30(11): 770-776.
[19]
Schmidt JC, Nietgen GW, Freisberg L, et al. Modified transscleral suture for sulcus fixation of posterior chamber lenses[J]. J Cataract Refract Surg, 2002, 28(1): 15-17.
[20]
Walkow L, Klemen UM. Patient satisfaction after implantation of diffractive designed multifocal intraocular lenses in dependence on objective parameters[J]. Graefes Arch Clin Exp Ophthalmol, 2001, 239(9): 683-687.
[21]
Hayashi K, Manabe SM, Hayashi H. Effect of astigmatism on visual acuity in eyes with a diffractive multifocal intraocular lens[J]. J Cataract Refract Surg, 2010, 36(8): 1323-1329.
[22]
Monaco G, Scialdone A. Long-term outcomes of limbal relaxing incisions during cataract surgery: aberrometric analysis[J]. Clinical Ophthalmology, 2015, 9(default): 1581-1587.
[23]
Toto L, Vecchiarino L, D'Ugo E, et al. Astigmatism correction with toric IOL: analysis of visual performance, position, and wavefront error[J]. J Refract Surg, 2013, 29(7): 476-483.
[24]
Venter JA, Pelouskova M, Collins BM, et al. Visual outcomes and patient satisfaction in 9366 eyes using a refractive segmented multifocal intraocular lens[J]. J Cataract Refract Surg, 2013, 39(10): 1477-1484.
[25]
Behndig A, Montan P, Stenevi U, et al. Aiming for emmetropia after cataract surgery: Swedish National Cataract Register study[J]. J Cataract Refract Surg, 2012, 38(7): 1181-1186.
[26]
Pierro L, Modorati G, Brancato R. Clinical variability in keratometry, ultrasound biometry measurements, and emmetropic intraocular lens power calculation[J]. J Cataract Refract Surg, 1991, 17(1): 91-94.
[27]
Raman S, Redmond R. Reasons for secondary surgical intervention after phacoemulsification with posterior chamber lens implantation[J]. J Cataract Refract Surg, 2003, 29(3): 513-517.
[28]
Snead MP, Rubinstein MP, Lea SH, et al. Calculated versus A-scan result for axial length using different types of ultrasound probe tip[J]. Eye, 1990, 4(5): 718-722.
[29]
Alio J, Plazapuche AB, Férnandezbuenaga R, et al. Multifocal Intraocular Lenses: An Overview[J]. Survey of Ophthalmology, 2017, 62(5): 611-634.
[30]
Pedrotti E, Carones F, Aiello F, et al. Comparative analysis of visual outcomes with 4 intraocular lenses: Monofocal, multifocal, and extended range of vision[J]. J Cataract Refract Surg, 2018, 44(2): 156-167.
[31]
Liao X, Lin J, Tian J, et al. Evaluation of Optical Quality: Ocular Scattering and Aberrations in Eyes Implanted with Diffractive Multifocal or Monofocal Intraocular Lenses[J]. Curr Eye Res, 2018, 43(6): 696-701.
[32]
Alió JL, Piñero DP, Plaza-Puche AB, et al. Visual outcomes and optical performance of a monofocal intraocular lens and a new-generation multifocal intraocular lens[J]. J Cataract Refract Surg, 2011, 37(2): 241-250.
[33]
Akondi V, Dorronsoro C, Gambra E, et al. Temporal multiplexing to simulate multifocal intraocular lenses: theoretical considerations[J]. Biomedical Optics Express, 2017, 8(7): 3410-3425.
[34]
De Vries NE, Laurendeau C, Lafuma A, et al. Lifetime costs and effectiveness of ReSTOR compared with a monofocal IOL and Array-SA40 in the Netherlands[J]. Eye, 2010, 24(4): 663-772.
[35]
Maxwell WA, Waycaster CR, D′Souza AO, et al. A United States cost-benefit comparison of an apodized, diffractive, presbyopia-correcting, multifocal intraocular lens and a conventional monofocal lens[J]. J Cataract Refract Surg, 2008, 34(11): 1855-1861.
[36]
Akaishi L, Vaz R. Visual Performance of Tecnis ZM900 Diffractive Multifocal IOL after 2500 Implants: A 3-Year Followup[J]. Journal of Ophthalmology, 2010, 2010(2090-004X): 717591.
[37]
Chen W, Meng Q, Ye H, et al. Reading ability and stereoacuity with combined implantation of refractive and diffractive multifocal intraocular lenses[J]. Acta Ophthalmologica, 2011, 89(4): 376-381.
[38]
Ji J, Huang X, Fan X, et al. Visual performance of Acrysof ReSTOR compared with a monofocal intraocular lens following implantation in cataract surgery[J]. Experimental & Therapeutic Medicine, 2013, 5(1): 277-281.
[39]
Alfonso JF, Puchades C, Fernández-Vega L, et al. Contrast sensitivity comparison between AcrySof ReSTOR and Acri.LISA aspheric intraocular lenses[J]. J Refract Surg, 2010, 26(7): 471-477.
[40]
Ravalico G, Parentin F, Sirotti P, et al. Analysis of light energy distribution by multifocal intraocular lenses through an experimental optical model[J]. J Cataract Refract Surg, 1998, 24(5): 647-652.
[41]
Martínez PA, Gómez FP, España AA, et al. Visual function with bilateral implantation of monofocal and multifocal intraocular lenses: a prospective, randomized, controlled clinical trial[J]. J Refract Surg, 2008, 24(3): 257-264.
[42]
Aslam TM, Gupta M, Gilmour D, et al. Long-term Prevalence of Pseudophakic Photic Phenomena[J]. Am J Ophthalmol, 2007, 143(3): 522-524.
[43]
Hofmann T, Zuberbuhler BA. Retinal straylight and complaint scores 18 months after implantation of the AcrySof monofocal and ReSTOR diffractive intraocular lenses[J]. J Refract Surg, 2009, 25(6): 485-492.
[44]
Donnenfeld ED, Solomon R, Roberts CW, et al. Cyclosporine 0.05% to improve visual outcomes after multifocal intraocular lens implantation[J]. J Cataract Refract Surg, 2010, 36(7): 1095-1100.
[45]
Cillino S, Casuccio A, Di PF, et al. One-year outcomes with new-generation multifocal intraocular lenses[J]. Ophthalmology, 2008, 115(9): 1508-1516.
[46]
Holladay JT, Dijk HV, Lang A, et al. Optical performance of multifocal intraocular lenses[J]. J Cataract Refract Surg, 1990, 16(4): 413-422.
[47]
Shoji N, Shimizu K. Clinical evaluation of a 5.5 mm three-zone refractive multifocal intraocular lens[J]. J Cataract Refract Surg, 1996, 22(8): 1097-1101.
[48]
MontãS-Micã R, Espaã AE, Bueno I, et al. Visual performance with multifocal intraocular lenses: mesopic contrast sensitivity under distance and near conditions[J]. Ophthalmology, 2004, 111(1): 85-96.
[49]
Rosen E, Alió JL, Dick HB, et al. Efficacy and safety of multifocal intraocular lenses following cataract and refractive lens exchange: Metaanalysis of peer-reviewed publications[J]. J Cataract Refract Surg, 2016, 42(2): 310-328.
[50]
Plaza-Puche AB, Alio JL, Sala E, et al. Impact of low mesopic contrast sensitivity outcomes in different types of modern multifocal intraocular lenses[J]. Euro J Ophthalmol, 2016, 26(6): 612-617.
[51]
Greenstein S. The Quest for Spectacle Independence: A Comparison of Multifocal Intraocular Lens Implants and Pseudophakic Monovision for Patients with Presbyopia[J]. Seminars in Ophthalmology, 2017, 32(1): 111-115.
[52]
Attia MS, Khoramnia R, Auffarth GU, et al. Near and intermediate visual and reading performance of patients with a multifocal apodized diffractive intraocular lens using an electronic reading desk[J]. J Cataract Refract Surg, 2016, 42(4): 582-590.
[53]
魏超,张晗. 多焦点人工晶状体的临床应用进展[J]. 山东大学耳鼻喉眼学报,2017,31(4):29-35.
[54]
Akondi V, Pérezmerino P, Martinezenriquez E, et al. Evaluation of the True Wavefront Aberrations in Eyes Implanted With a Rotationally Asymmetric Multifocal Intraocular Lens[J]. J Refract Surg, 2017, 33(4): 257-265.
[55]
Mckee HD, Jhanji V. Theoretical Effect of Lens Position and Corneal Curvature on the Near Focal Point of Multifocal Intraocular Lenses[J]. J Refract Surg, 2016, 32(1): 64-66.
[1] 马丹, 李雅楠, 张丽, 苗金红, 李学民, 胡晋平. 不同人工晶状体装载方法对术中襻与光学部黏附影响的临床研究[J]. 中华眼科医学杂志(电子版), 2023, 13(01): 24-29.
[2] 万修华. 角膜移植术后白内障吸除联合张力环及后房型人工晶状体植入术[J]. 中华眼科医学杂志(电子版), 2022, 12(06): 0-.
[3] 万修华. 白内障标准化手术系列视频(5级全白核)[J]. 中华眼科医学杂志(电子版), 2022, 12(05): 0-.
[4] 王晓宇, 李亚新, 刘一昀, 耿嘉懿, 秦锐, 李炎城, 敖明昕, 刘德海, 齐虹. 不同设计多焦点人工晶状体植入后视觉质量差异的临床研究[J]. 中华眼科医学杂志(电子版), 2022, 12(04): 210-215.
[5] 姜雅琴, 刘丽峰, 刘秀花, 张亚丽. 白内障术后早期患者配戴绷带镜对泪膜质量及其动态变化影响的临床研究[J]. 中华眼科医学杂志(电子版), 2022, 12(04): 204-209.
[6] 万修华. 白内障标准化手术系列视频(4级核)[J]. 中华眼科医学杂志(电子版), 2022, 12(03): 0-.
[7] 熊天旭, 范玮, 万修华. 白内障患者老视矫正的研究进展[J]. 中华眼科医学杂志(电子版), 2022, 12(01): 52-56.
[8] 万修华. 白内障标准化手术系列视频(2级核)(视频)[J]. 中华眼科医学杂志(电子版), 2022, 12(01): 0-0.
[9] 姚沁楠, 万修华. 有晶状体眼后房型人工晶状体植入术与角膜屈光手术治疗高度近视眼有效性、安全性及可预测性的Meta分析[J]. 中华眼科医学杂志(电子版), 2021, 11(06): 346-352.
[10] 王珏雪, 王开杰, 万修华. 人工晶状体相关并发症的研究进展[J]. 中华眼科医学杂志(电子版), 2021, 11(05): 311-315.
[11] 李猛, 王进达, 张景尚, 王开杰, 毛迎燕, 陈淑莹, 姚沁楠, 万修华. 六种人工晶状体计算公式计算先天性白内障患者人工晶状体度数准确性的临床研究[J]. 中华眼科医学杂志(电子版), 2021, 11(04): 211-216.
[12] 刘兆川, 宋旭东. 重视景深延长型人工晶状体在屈光性白内障手术中的应用[J]. 中华眼科医学杂志(电子版), 2021, 11(04): 193-197.
[13] 陇悦, 范玮. 白内障手术中复曲面人工晶状体拓展运用的临床现状[J]. 中华眼科医学杂志(电子版), 2021, 11(02): 109-113.
[14] 姜雅琴, 张敏, 马健利, 黄旭东, 宋丽平. 飞秒激光辅助超声乳化术后不同人工晶状体有效位置和轴向位移量变化的临床研究[J]. 中华眼科医学杂志(电子版), 2021, 11(02): 76-81.
[15] 何海龙, 刘振宇, 周春媛, 张立平, 王进达, 万修华. 飞秒激光小切口角膜基质透镜切除术与有晶状体眼后房型人工晶状体植入术矫正高度近视眼疗效的Meta分析[J]. 中华眼科医学杂志(电子版), 2021, 11(01): 22-28.
阅读次数
全文


摘要